Search results for "Ring down"
showing 2 items of 2 documents
Acetone measurements in the exhaled air by the cavity ring-down spectrometry
2020
Our work is focused to develop sensitive optical diagnostic methods for the detection of Volatile Organic Compounds in the breath. For the breath measurements, we have built a cavity ring-down spectrometer, operated in the UV region. The system was tested and calibrated using a KinTek automated permeation tube system with nitrogen as a carrier gas, as well as self- prepared acetone mixtures in air with known concentrations. In this paper, we present results of the measurements of acetone concentration in the breath for the test group of randomly selected 40 people in the age group over 50. We observed the values of acetone concentration in the breath in the range from 0,1 – 80 ppm, with a m…
First results of cavity ring down signals from exhaled air
2017
In this paper we report first results from the developed cavity ring-down spectrometer for application in human breath analysis for the diagnostics of diabetes and later for early detection of lung cancer. Our cavity ring-down spectrometer works in UV region with pulsed Nd:YAG laser at 266 nm wavelength. First experiments allow us to determine acetone and benzene at the level bellow ppm. In our experiment, first results from breath samples from volunteers after doing different activities were collected and examined. Influence of the smoking on the breath signals also was examined.